MXU Motor Drive Board Repair Instructions
print icon
  1. Purpose
    1. This work instruction describes the process for replacing the MXU Motor Drive Board.
    2. These repair instructions are valid for revision C MXU Digital Centrifuges. For revisions older than C, please consult Product Development.
  2. Equipment/Software
    1. ITEM # PART # DESCRIPTION QTY.
      1 100294 Vibration Absorber; Rubber 3
      2 101348 Machine screw; M6 3
      3 100067 Washer; split lock; M6 6
      4 100057 Nut; hex; M6 3
      5 100952-E Base Plate; MX12 1
    2. Needle Nose Pliers
       
  3. Instructions
    1. Confirm that all necessary components are present to complete the conversion.
    2. Open the lid and remove the rotor and tube shields from the unit. Figure 1.


       
    3. Turn the unit over onto its lid and remove the base plate. Figure 2.
      NOTE: The motor wiring harnesses will be connected. Disconnect them or set the base close enough to the side to not overextend the wires.


       
    4. Locate Motor Drive Board on front face of unit. Figure 3.


       
    5. Remove all connected wiring harnesses from Motor Drive Board.
    6. Using needle nose pliers, compress the standoffs that are holding the board in place. Figure 4.

    7. Remove the non-working Motor Drive Board (100480) and replace it with a new one. Be sure to snap-in all 4 standoffs. Figure 5.


       
    8. Re-connect all cables as shown. Figure 6.


       
    9. Re-attach the base plate. Figure 7.


       
    10. Replace Rotor, close the lid, and verify the unit powers on and functions under normal laboratory use.


       

      SNR-7.5.1-W-005 | Rev 0