MXU Motor Drive Board Repair Instructions

* Purpose This work instruction describes the process for replacing the MXU Motor Drive Board. These repair instructions are valid for revision C MXU Digital Centrifuges. For revisions older than C, please consult Product Development. * Equipment/Software ITEM # PART # DESCRIPTION QTY. …

MX12 Lid Latch Repair

* Purpose This work instruction describes the process for replacing the lid latch on MX12 centrifuge units. * Equipment/Software Replacement Lid Latch (101078-D) Attachment Screws, M4 x 8mm, button head hex drive, stainless steel 2.5mm hex driver * Instructions * Check that you have all necessary c…

Service Light and Display Board Reset Instructions E8D Universal

Service Light Reset Instructions E8D/Universal With the power on, begin by closing and latching the lid, you should have a steady green light on the front panel. * Set TIME display to 00 * Set SPEED display to 00 * Press the STOP button At this point the two displays should be flashing dashes ( --  --) in …